Define the boundary, time basis and measurable product form

A perfusion balance asks where liquid, cells and product go during a defined time interval. The general structure is identical for each quantity: inventory change = input − output + formation − degradation. The transported components differ. Cell-free medium can add volume and substrate without introducing viable cells. A cell-retention filter can remove liquid and dissolved product while returning most cells. A bleed deliberately removes culture, including cells.

Mark the system boundary on the flowsheet before calculating. If it encloses the reactor, retention module and loop together, recirculation is internal. If it encloses only the reactor, the two circulation connections become boundary streams. Both approaches are possible. Combining them creates double-counted flows or disappearing inventories. The same applies to a harvest line, cooler or collection bag: its inventory is either inside or outside the chosen boundary.

Define product analytically. Total protein, intact target protein, monomeric target and biological activity are not interchangeable. Cleavage of an intact protein represents loss of that product form even though total protein mass need not disappear. An apparent balance residual can therefore represent an unmeasured analytical fraction. It is not automatically evidence of measurement error or biological degradation.

All examples below are explicitly synthetic. They describe an extracellular dissolved product in ideally mixed liquid. A day is 24 h; volumes are litres, product masses are grams and cell numbers refer to viable cells. A real culture additionally needs evidence for volume measurement, evaporation, dosing, samples, flushes and representative concentrations.

Close the liquid balance first

Let Ff denote fresh medium, Fa other additions, Fh permeate harvest, Fb bleed and Fs sampling. With liquid volumes approximated as additive, dV/dt = Ff + Fa − Fh − Fb − Fs − Fev. Here Fev is net evaporative loss expressed as liquid volume. Condensate return must follow its actual route. Gas volume is not simply added to liquid volume.

In the teaching case, V = 100 L, Fh = 80 L/d, Fb = 5 L/d and Fs = 0.2 L/d. Other additions and evaporation are set to zero. Constant inventory therefore requires 85.2 L/d of fresh medium. Replacing only the 80 L/d harvest would lose 5.2 L each day, or 31.2 L over six days. This is not a rounding difference. A level controller might conceal the error by introducing an unrecorded make-up stream.

Sampling is represented by an equivalent continuous flow in this model. Many real samples are discrete events. At each withdrawal, subtract ΔV and remove c × ΔV of product mass. Replacing the sample with product-free medium may also change the subsequent concentration. A continuous approximation can sometimes represent small frequent withdrawals; larger events need explicit assessment.

Converting a gravimetric balance to volume requires appropriate density. Fresh medium, concentrated feed and harvest may have different densities. Tubing forces, tare changes and container replacement can also create apparent flows. Weight loss from a supply vessel demonstrates departure from that vessel, not necessarily arrival at the reactor. Feed mass-to-volume conversion and a documented line inventory close that gap.

Connect viable-cell inventory, bleed and growth

Viable-cell inventory is Nv = XvV. With cell-free fresh medium and well-mixed bleed, dNv/dt = (μ − kd)Nv − FbXv − FsXv − FhXh. Here μ represents specific formation of viable cells and kd their loss through the modelled death processes; both are in d⁻¹. The equation distinguishes biological loss from physical removal. A bleed is not cell death.

At constant volume and Xv, μ − kd = (Fb + Fs + FhsX)/V. The assumed cell passage sX = 0.00025 gives a required net cell formation rate of 0.0522 d⁻¹. If a death rate of 0.02 d⁻¹ were additionally assumed, μ would be 0.0722 d⁻¹. None of these numbers was estimated from a real CHO culture. They demonstrate which rates consistently sustain the assumed constant cell inventory.

Fresh-medium CSPR is explicitly defined here as Ff/Nv. With 40 × 10⁹ viable cells/L and 100 L, Nv is 4 × 10¹² cells and CSPR is 21.3 pL/(cell·d). A quantity based only on permeate flow would be 20.0 pL/(cell·d). Both calculations are correct but answer different questions. Reports must identify which liquid stream is included in the numerator.

Constant measured cell concentration does not prove stationary cellular behaviour. Growth and loss may simply balance, while different cell states can produce the same count. Increasing volume can also reduce concentration while total cell inventory grows. With varying volume, balance Nv first. Productivity is subsequently normalised to integrated viable-cell time, rather than an arbitrarily selected final concentration.

Derive the product balance with changing sieving

For product inventory M = cV, product-free fresh medium and specific formation rate qP, dM/dt = qPNv − FhsPc − (Fb + Fs)c − kPM. Here qP is in g/(cell·d), and kP is a degradation constant in d⁻¹ chosen solely for the teaching case. Filter adsorption, additional liquid inventories or other product forms would require separate terms. They are not silently included in bleed.

With constant V and constant conditions within an interval, dc/dt = G/V − ac, where G = qPNv and a = (FhsP + Fb + Fs)/V + kP. Both terms on the right have units of g/(L·d). For a > 0, ceq = G/(Va) is the concentration towards which the model would move if those interval conditions continued. It is not an automatically attained daily result.

Solving the linear differential equation gives c(t+Δt) = ceq + [c(t) − ceq] exp(−aΔt). The concentration integral is Ic = ceqΔt + [c(t) − ceq][1 − exp(−aΔt)]/a. Integrated harvest mass is FhsPIc, with analogous expressions for bleed, sampling and degradation. This closed-form solution enables an exact balance check for the synthetic case, without confusing numerical integration error with missing material.

If sP falls, effective product removal decreases. Concentration can consequently rise at unchanged formation, while harvest delivery falls. These observations are consistent: more product remains inside the system. Whether it can later be recovered at acceptable quality is a separate question. The definition of observed sieving is therefore central to interpreting a high reactor titre.

Reconcile six complete days

The case begins at 1.0 g/L, corresponding to 100 g of product inventory. Viable-cell number remains at 4 × 10¹². With qP = 20 pg/(cell·d), formation G is 80 g/d, because one picogram is 10⁻¹² g. Exactly 480 g is therefore formed over six days. Assumed product degradation is kP = 0.01 d⁻¹. Product sieving changes stepwise at daily boundaries and remains constant within each day.

The table gives each day's final concentration, integrated harvest and inventory change. Harvest is not daily final concentration multiplied by daily volume; it integrates the concentration actually represented throughout that day. Each interval starts from the preceding interval's final condition.

After six days, integrated permeate harvest is 348.1038 g. Bleed removes 36.3997 g, continuously approximated sampling removes 1.4560 g, and assumed degradation removes 7.2799 g. Inventory increases by 86.7605 g to 186.7605 g. The balance therefore closes: 480.0000 = 348.1038 + 36.3997 + 1.4560 + 7.2799 + 86.7605 g, apart from displayed rounding.

Including starting inventory gives the equivalent overall view: 100 + 480 g of starting material and formation = 348.1038 g harvest + 45.1357 g other losses + 186.7605 g final inventory. Starting material must not be forgotten when terminal reactor contents are considered potential harvest. Equally, final inventory must not be counted both as already harvested and as still present in the reactor.

Synthetische Intervallbilanz / Synthetic interval balance
Intervall / intervalsPc Ende / end (g/L)Ernte / harvest (g)Δ Bestand / inventory (g)
0–10.950.985075.3528-1.5000
1–20.91.005671.75832.0625
2–30.81.073266.77356.7578
3–40.61.241755.922416.8543
4–50.41.512444.341727.0671
5–60.251.867633.955135.5188
Product formation differs from product harvest. Daily model formation, harvest and signed inventory change. Bleed, sampling and degradation are additional terms in the chapter’s complete balance.
Daily model formation, harvest and signed inventory change. Bleed, sampling and degradation are additional terms in the chapter’s complete balance. Scroll sideways on small screens or open to enlarge.

Integrate measurements without pretending to know more

Real time series usually do not provide the closed-form model solution. If flow F and harvest concentration ch are known at sampling times, harvest rate rh = Fch can, for example, be linearly interpolated and integrated with the trapezoidal rule. The resulting mass depends on sample spacing, signal alignment and interpolation assumptions. A pump recorded every second does not create second-by-second knowledge of titre measured only once daily.

With changing flow and concentration, the average of their product is generally not the product of their averages. Consider two equal intervals: first 2 L/h at 1 g/L, then 1 L/h at 2 g/L. One hour at each condition yields 4 g in total. Multiplying mean flow, 1.5 L/h, by mean titre, 1.5 g/L, and two hours would give 4.5 g. Time alignment is therefore part of mass measurement.

A collection bag can provide a useful time-integrated measurement. Its total product mass follows from a representative mixed concentration and the actual contained volume. Residual inventory, sampling, temperature or density basis, and adequate mixing must still be known. An unweighted mean of time samples is not automatically equivalent to a flow-proportional composite sample.

Report data gaps explicitly. Plausible interpolation can support a working estimate, but is not a retrospective measurement. For an important gap, calculate a defensible range using plausible bounds and examine whether the decision depends on it. An unknown mass should not be made to disappear by freely choosing a degradation coefficient. A balance residual is initially an observation requiring investigation.

Terminal harvest and stopping change the boundary

If part of the final reactor inventory is recovered, report terminal harvest separately. A purely hypothetical terminal recovery of 90% would obtain another 168.0845 g from the 186.7605 g inventory. Together with previously collected permeate harvest, this gives 516.1883 g. Exceeding the 480 g newly formed during the observation window does not violate the balance: 100 g was present initially. Comparing cultures therefore requires matching initial and final conditions or explicit inventory correction.

The assumed terminal-harvest loss is 18.6761 g. It is additional to losses during the six operating days and must not also be counted as culture degradation. A subsequent capture yield would represent another step, referenced to its actual inlet. If permeate and terminal harvest are processed separately, their yields and quality attributes may differ. A single blanket percentage would conceal those differences.

After an unexpected interruption, accounting does not end with the last normal data point. A fault may cause discharge, dilution, holding, degradation or restriction of previously collected material. Physically present mass, analytically suitable mass and mass accepted for subsequent use then become three different reporting levels. The physical balance should still close: discarded material is a documented outlet, not disappearing mass.

Campaign-based assessment connects these quantities to occupied time and medium consumption. That is where the appropriate recovery and denominator for a specific engineering question are selected.

Do not infer formation and degradation from one titre

The product balance contains both formation and degradation. An observed concentration profile may determine only their net effect. At a simple steady state, G = c(FhsP + Fb + Fs + kPV). Without independent information on kP, different combinations of formation and degradation can give the same concentration. Even a perfectly fitted curve therefore does not prove that both biological quantities have been identified uniquely.

For an independent numerical check, take c = 1 g/L, V = 100 L and product-effective outlet flow FhsP + Fb + Fs = 80 L/d. With no degradation, G = 80 g/d fits. With kP = 0.05 d⁻¹, G = 85 g/d fits the same steady titre. Additional formation replaces exactly five grams per day of assumed degradation. Concentration alone cannot distinguish the explanations.

A suitable additional observation could establish stability of the actual measured product form under relevant conditions. Further fraction and loss measurements may also constrain the balance. The useful observation separates explanations previously coupled together. Another precise titre in the same stationary state may contribute less than an independent, appropriately bounded stability experiment. Report estimated net formation differently from individually supported rates.

Balance mass first when volume changes

With changing volume, d(cV)/dt = V dc/dt + c dV/dt. A product-mass equation therefore cannot be used as a concentration equation while omitting the second term. Start from inventory M and derive concentration afterwards. This prevents dilution-driven concentration changes from being misinterpreted as biological degradation.

A simple example starts with 100 L at 1 g/L, or 100 g of product. Ten litres of product-free liquid enter over one hour; formation, degradation and outflow are zero. The final state is 110 L with the same 100 g, giving 0.9091 g/L. The approximately 9.09 per cent concentration decrease is entirely explained by volume. An analysis that continues assuming 100 L would report a fictitious loss of approximately 9.09 g.

When incoming liquid itself contains product, Fincin becomes a separate mass input. An analytical reference material introduced with a liquid, for example, is not automatically newly formed culture product. Concentrated additions likewise need their actual quantities included in liquid and component balances. Additive liquid volumes may be a useful approximation in some calculations; where density or mixing changes matter, a mass balance using measured density is clearer.

Use residuals diagnostically rather than as hidden corrections

An actual balance rarely closes to the last displayed digit. Define its residual with an explicit sign, for example observed inventory change minus the change calculated from inputs, outputs and assumed reactions. A positive residual then means something different from the reverse definition. Include the convention before discussing causes.

Residuals can arise from missing streams, incorrect timing, measurement uncertainty or an unsuitable reaction model. A small difference between two large inventories is particularly sensitive to shared or independent measurement errors. At 100 L and approximately 1 g/L, inventory is around 100 g. A concentration difference of 0.01 g/L already represents approximately 1 g; a daily increase of only a few grams therefore needs suitable uncertainty consideration.

Freely adjusting formation rate afterwards can remove the residual mathematically without explaining its cause. Report the raw balance, justified corrections and remaining residual separately. Reconstructed missing samples must retain visible estimation assumptions. The exactly closing synthetic case primarily checks calculation consistency; it does not promise equivalent closure or biological-rate identifiability in measured data.

Exercises with checkable answers

Check 1: Bleed increases from 5 to 8 L/d with other conditions unchanged. What must initially change to maintain 100 L total volume? Fresh-medium flow increases from 85.2 to 88.2 L/d unless another stream compensates. The net viable-cell formation rate required for constant inventory increases by 3/100 = 0.03 d⁻¹. This does not mean the cells will actually grow faster. Without that biological response, cell inventory decreases.

Check 2: Sieving and flow remain constant while reactor titre doubles. What follows immediately? Instantaneous product harvest rate FhsPc doubles. Cumulative daily harvest doubles only if the corresponding time course doubles throughout the entire interval. One final point is insufficient.

Check 3: A reactor sample contains 2 g/L. Of 0.5 L withdrawn, 0.1 L is discarded before analysis. How much product leaves the reactor? With representative withdrawal, a total of 1 g leaves, regardless of whether only 0.4 L is analysed or retained. Accounting only for analytical volume omits 0.2 g. Reactor sampling and laboratory disposition are different boundaries.

Check 4: The model closes within 10⁻¹⁰ g. Does this mean a real process can be balanced that accurately? No. The test checks mathematical consistency of chosen inputs and their implementation. Empirical closure requires measurement uncertainty, representative samples and complete flow accounting. A numerically precise model can still make an incorrect real-world statement if its assumptions are unsuitable.

Data, calculation and original sources

The synthetic interval values, Python calculation and machine-readable results form one package. The calculation uses only the Python standard library. Its data dictionary explains the boundary, units and chosen assumptions.

Repligen's description of perfusion culture documents the basic arrangement of exchange and retention. The balances developed here explicitly distinguish permeate, bleed and samples. Commercial productivity comparisons are not adopted.

The original study by Liang and colleagues illustrates why membrane choice and product location must be described. The dynamic case here is independently derived. It reproduces neither their culture nor a validated production process. Continue with productivity and harvest planning.