Turn the process profile into testable requirements
Start with a short process profile: organism or cell type, medium, product, required amount, operating mode and culture duration. Add expected cell concentration, oxygen demand, viscosity, foaming and any particular requirements for light or carrier particles. For a development process, these can be uncertain ranges rather than apparently precise single values.
Translate the profile into requirements that can be checked. Instead of stating that a system must be suitable for demanding fermentation, specify the oxygen transfer capability needed in the proposed medium and working volume. A development reactor may need broad adaptability, whereas a production process may require repeatable operations and sufficient reserve within a narrower operating window.
A useful profile starts with the outcome: what quantity, with which properties, must reach the next step, and when? Working volume and biological load follow only after assumptions about concentration, yield and harvest loss. For an unfamiliar process, carry these assumptions as ranges. A single expected final concentration should not simultaneously determine design, capacity planning and acceptance when its variability has not yet been investigated. Procurement needs to distinguish requirements based on measurements from those introduced provisionally.
The user requirements specification, often called the URS, translates that task into verifiable statements. Give each important requirement a unique identifier, rationale, operating conditions, measured quantity, acceptance criterion and intended evidence. Assign responsibility for unresolved information as well. “Good cooling” then becomes a stated heat load at a defined fill, culture temperature, available cooling-water supply and permissible deviation. The figures come from the process and site; this chapter does not provide universal limits.
Distinguish mandatory criteria, scored preferences and development questions. A mandatory criterion might be a particular minimum fill with effective measurement. A preference might be an additional spare interface. A development question might concern gas transfer in a viscous culture expected later. Marking that open question as satisfied creates false certainty. Treating it as necessarily impossible without investigation may prematurely remove a useful option. The URS should therefore show how and when each open question is to be resolved.
| Initial wish | Necessary clarification | Suitable evidence |
|---|---|---|
| Wide volume range | Initial and final fill, with functions required at both | Geometry review and functional trials |
| High oxygen supply | Demand, medium, gas, pressure and permitted actuators | Transfer characterisation under stated conditions |
| Accurate dosing | Minimum and maximum flow, duration and permissible error | Gravimetric trial with the intended tubing path |
| Complete data | Signals, time basis, events and usable export | Record, export and reopen a test run |
| Easy changeover | Actual sequence, personnel and available time | Walk-through with intended components and operator roles |
A requirement preferably describes the needed function. Specify a particular technical solution when connections, existing processes or justified compatibility make it necessary. Otherwise, a narrow early preference may exclude suitable alternatives. Ask whether another construction could demonstrate the same required outcome under identical conditions. If it could, document the function separately from the implementation currently preferred.
Consider working volume throughout the process
Depending on the supplier, nominal volume may refer to total vessel capacity or a specified working volume. Total capacity must be distinguished from usable working volume. Space must remain for headspace, foam, internals and a reliable exhaust route. Equally, the reactor needs a minimum fill for sensors to remain immersed, impellers to operate effectively and temperature control to work adequately.
A fed-batch process must operate both after inoculation and at the end of feeding. Perfusion also requires consideration of the liquid held in the retention device, tubing and any intermediate vessels. That volume cannot be inferred from vessel capacity alone.
A useful specification might state: start at 2 L and finish at 5 L, with a defined medium and proposed gas supply. This is a requirement that can be checked. Asking for a 5 L bioreactor leaves open whether 5 L describes total capacity or usable culture liquid.
Calculate working volume as a trajectory. Liquid additions include nutrient feed, correction liquids, antifoam and any rinsing or transfer quantities. Outlets include samples, bleed, harvest and water losses. Geometric assessment additionally considers gas hold-up and foam, which can change occupied space. Calculated liquid volume is therefore not identical to the highest observed liquid or foam level during aeration.
Hypothetical teaching example: A process starts with 2.0 L and receives 2.4 L of nutrient feed and 0.3 L of other liquids before harvest. Total sampling is 0.2 L and assumed water loss is 0.1 L. Calculated final volume is 2.0 + 2.4 + 0.3 − 0.2 − 0.1 = 4.4 L. Under simplified volume additivity, the configuration must function across at least the 2.0–4.4 L trajectory. This does not assume that overflows, unplanned losses or gas entrainment are acceptable.
The ratio of greatest to smallest intended liquid fill is 4.4/2.0 = 2.2. This volume turndown describes a requirement only. It does not establish that every pump, probe and heat transfer surface remains useful across the same factor. At small fill, relevant internals must be wetted and additions distributed appropriately; at large fill, gas transfer, headspace and heat removal must suffice. Selecting equipment by final volume can easily overlook the initial condition.
For an additional hypothetical perfusion option, assume 0.4 L in the module and tubing. At 2.0 L main-vessel fill, the connected liquid system therefore contains 2.4 L. External exchange of 2.4 L/d is 1.0 d−1 on the complete-system basis, but 1.2 d−1 on the main-vessel basis. Both calculations can be correct while answering different reference questions. Explicitly identify the denominator in the requirement and the report.
Flow ranges also need separate assessment. An assumed feed range of 2–200 mL/h requires a factor of 100 between minimum and maximum commanded delivery. Measure whether the dosing assembly achieves this over the required duration and within the permitted error. Different tubing sizes or separate dosing routes may be useful; their assembly and changeover then become part of operation and aseptic design. A large maximum pump capacity does not answer the question of the smallest reliable addition.
Glass: visibility with practical limitations
Glass vessels allow direct observation of foam, liquid level and culture movement. Reusable vessels with metal headplates and interchangeable internals are common in laboratories. Being able to dismantle the assembly makes modifications easier, but also introduces cleaning, assembly and seal-checking work.
Autoclave compatibility must cover the complete loaded vessel, including sensors, filters and tubing. Weight, dimensions, transport route and available autoclave chamber are real selection constraints. Temperature changes and mechanical handling must remain compatible with the glass and the manufacturer's instructions. Transparency is no evidence of pressure capability; permitted positive and negative pressures require separate confirmation.
Visibility is particularly useful when a development process requires observation of foam, aggregates or changing carrier distribution. It supplements measurements rather than replacing them. A visible vortex does not describe a defined mixing time, and an uneventful appearance does not establish culture purity. State in the URS why optical access is needed: observation, illumination or an optical measurement. Each purpose may create different requirements for surface, geometry and position.
A hypothetical site comparison shows why the surroundings matter. An existing autoclave might accommodate the bare vessel but not the complete assembly with filters, probes and protected tubing. Offered vessel capacity alone is then insufficient evidence of compatibility. Assess the real loaded configuration, transport route and planned handling. If process-essential components repeatedly require a different arrangement for sterilisation, include the extra operation in workload and error assessment.
For fair comparison, distinguish reusable contact parts, wear parts and necessary spare assemblies. An additional vessel assembly may separate preparation from cultivation in time, but it also increases the components needing cleaning, storage and identification. Whether this is worthwhile depends on the intended sequence. Material selection should therefore end with a traceable walk-through of the actual work cycle, rather than a general judgement about laboratory size or flexibility.
Stainless steel: permanent integration needs proper design
Stainless steel offers many options for robust vessels, permanent piping and effective heat exchange. Suitable systems can be cleaned and sterilised in their installed position. Cleaning in Place, or CIP, describes cleaning without the dismantling otherwise needed for that task. Sterilisation in Place, or SIP, sterilises the specified system where it is used.
These capabilities do not follow from the material alone. Geometry, drainage points, valves, seals and internals must support the intended procedure. Cleaning supply, steam, cooling and effective condensate handling require space and utilities. Stainless steel also has limits to chemical resistance. Material grade, surface condition and cleaning chemicals must be compatible with the process.
The benefits of permanent integration should be weighed against modification work and potential downtime during product changeover.
For a permanently installed system, scope boundaries become particularly important. A supplied cleaning device may assume a certain pressure and flow at its connection; the building must provide those conditions while other consumers operate simultaneously. A pipe present in the room does not establish that capability. Return flow, drainage, condensate and removal of released heat need consideration as well. Supply and disposal together provide the usable function.
A cleaning requirement describes the intended contact path and residues to be removed. Medium residues, cellular material and deposits can create different questions from a readily soluble test substance. Design review considers access, drainage and the surfaces covered; later evidence must suit the actual cleaning objective. Material designation alone establishes neither successful cleaning nor suitable sterilisation. Agree the associated procedures and their limits as part of the system.
An assumed multiproduct operation illustrates the connection. Fixed piping may simplify recurring work, while frequently changing additions require modification. Walk through a representative product change before deciding: which parts change, which checks follow, and how long the installation remains occupied. A planned reserve of adaptable connections may help. An unlimited number of unused ports is not an objective in itself, because every additional contact point enters assembly, cleaning and documentation.
Single-use: replace the product-contact flow path
In a single-use system, a specified product-contact flow path is replaced after use. The drive, support, temperature-control equipment and controller can remain in service. Single-use therefore does not mean that the entire installation is discarded. Systems may use flexible bags or rigid polymer vessels, and their mixing and aeration principles are not automatically identical.
A prepared, sterile-supplied flow path can simplify changeover. Its integrity still needs to survive transport, installation and operation. Material compatibility, connections and permitted loading require a specific assessment.
Extractables are substances released under selected extraction conditions; leachables migrate into process material under actual use conditions. Supplier data must fit the application. Single-use is consequently neither a universal sterility promise nor a substitute for process assessment. Long-term availability, lead times and change notifications for consumables provide further selection criteria.
Selection assesses the precise bill of materials for the contact path. A bag alone may omit required feed lines, sensors, retention modules or harvest containers. A meaningful comparison defines the complete assembly, including components connected later. Material information, integrity, permissible operating conditions and per-run costs refer to that configuration. An inexpensive base consumable says little about the effort required for the actual process assembly.
Contact duration, temperature, medium composition and subsequent processing influence material assessment. Review which of these conditions the available supplier data cover. Extractables investigations differ from determining which leachables actually enter the product. Product can also bind to contact surfaces without visible damage. The studies needed depend on product, application and process risk; research and regulated manufacturing do not automatically require identical evidence.
A hypothetical long-running process with several planned connection changes needs a different assessment from a short closed run. Consider integrity throughout the intended duration, permissible tensile loading and pressure conditions, and practical execution of connections. Detailed handling follows compatible components and documented procedures. The decision should establish whether the planned manipulations are represented in the proposed assembly, rather than simply whether one bag is supplied sterile.
Supply of suitable consumables is another operating function. Required information includes configuration identity, storage conditions, shelf life, ordering and change information, and justified inventory planning. An alternative source is a genuine reserve only if material, connections and process performance are sufficiently comparable. A visually similar item is not automatically interchangeable. Keep the assessment concrete: what would a contact-path change mean technically and in time, and which parts of the evidence would it affect?
Compare material systems using the same questions
| Question | Reusable glass | Reusable stainless steel | Single-use |
|---|---|---|---|
| Preparation | Cleaning, assembly and a suitable sterilisation method | Equipment cleaning and sterilisation procedure | Provide, install and aseptically connect the flow path |
| Process changes | Often possible through interchangeable internals | Depends on ports and permanent piping | Depends on available flow-path configurations |
| Operating resources | Autoclave capacity for autoclavable designs | Utilities matching CIP/SIP and process requirements | Consumables, storage and a disposal route |
| Performance evidence | Demonstrate mixing, gas transfer, cooling and measurement across the intended process range | ||
A numerical score becomes useful only after exclusion criteria have been checked. A system that cannot meet a required minimum volume or chemical resistance does not become suitable because its price is attractive. Weightings should reflect the actual process and should not create unsupported supplier rankings.
Hypothetical matrix example: Three invented configurations, A, B and C, are evaluated against identical requirements. They represent neither manufacturers nor material classes. Mandatory criteria precede scoring: functional initial fill of 2.0 L, a suitable contact path and the required data handover. In this example, A and B have sufficient evidence. C cannot provide the necessary functions at the required initial fill and is excluded. Other favourable properties must not compensate numerically for this failure.
| Assessment area | Weight | A | B | C, excluded |
|---|---|---|---|---|
| Demonstrated transport reserve | 35% | 4 | 3 | 5 |
| Volume and dosing adaptability | 20% | 4 | 5 | 2 |
| Fit with the work sequence | 20% | 3 | 4 | 4 |
| Usability of data | 15% | 5 | 3 | 4 |
| Supply and maintainability | 10% | 4 | 3 | 4 |
The weighted sum is S = Σ wi · si, where dimensionless weights wi sum to one and si are the scores. For A, 0.35 · 4 + 0.20 · 4 + 0.20 · 3 + 0.15 · 5 + 0.10 · 4 = 3.95. B scores 3.60. C would mathematically score 3.95 but remains excluded. An unresolved mandatory requirement remains open until demonstrated; it is not replaced by an intermediate numerical score.
Define score meanings before comparison. “Five points” must not simply mean the most persuasive conversation. Transport reserve might use agreed levels of evidence and process-specific margins; operation might use an observed sequence with identical tasks. Reserve earns credit only when usable under the jointly required conditions. Avoid accidentally weighting the same property repeatedly, for example by counting the full benefit of short changeover time separately in operation, availability and cost.
The decision depends on the process profile. In a second hypothetical case, A and B remain technically suitable, but frequent changes and a wide volume range become more important. Weights of 10% for transport, 25% for adaptability, 40% for work sequence, 10% for data and 15% for supply give A = 3.70 and B = 3.90. B now leads. This is the consequence of different priorities, not a contradiction. It would, however, be inappropriate to change weights after viewing the results merely to make a preferred alternative win.
A sensitivity assessment asks which defensible weights and still-uncertain scores would alter the ranking. When candidates are close, the appropriate next step is often a targeted investigation of the decisive uncertainty. Two decimal places do not make subjective scoring precise. The matrix structures discussion and records reasons; it replaces neither mandatory checks nor engineering judgement.
Assess functions as complete connected systems
Agitation performance requires appropriate impellers and their operating range. Aeration includes gas sources, controlled flows, spargers and the exhaust route. Temperature control depends on heat generation, heat exchange area and available cooling conditions. Each connected system is constrained by its limiting component.
Sensor ports must support the required measurements at minimum working volume. Pumps need sufficient resolution at the lowest feed rate and enough capacity at the highest. Additional probes, sampling and perfusion require consideration of spare connections, weights, tubing layout and eventual operator access.
A specification sheet may quote maximum values that apply at different operating points. Confirm whether the required agitation, gas flow, temperature and volume can be achieved simultaneously. A kLa value measured in water must be treated as a water result; it does not demonstrate identical performance in a viscous culture medium.
Build the gas requirement for each necessary route and major process phase. Specify composition, minimum and maximum flow, reference conditions, pressure required at the connection and intended simultaneous demand. A flow controller with a large maximum range may have a different relative error at very small flows. Display resolution is not the same as accuracy of delivered flow. During gas changes, line volume, switching logic and time until the vessel actually receives the new composition also matter.
Hypothetical consumption example: A defined process uses 6 h at 3 L/min followed by 18 h at 8 L/min of the same dry reference gas. Consumption is 6 · 60 · 3 + 18 · 60 · 8 = 9720 L, or 9.72 m³ at those reference conditions. Purging, preparation and additional consumers are not yet included. Peak flow remains 8 L/min; designing supply only around the average of 6.75 L/min would not fully describe the load case.
Parallel vessels require a justified simultaneous-demand assumption. If two systems each require 8 L/min in the most demanding planned concurrent state, their shared supply must support that state and agreed losses. For inventory, add the complete operating profiles instead. Storage capacity and instantaneous delivery capability are therefore different demonstrations. With compressed-gas containers, usable inventory follows from actual conditions and the permissible residual state; a general rule of thumb does not replace that assessment.
For cooling, assess heat load and heat transfer capability separately. If the hypothetical process requires removal of 1.4 kW and cooling water may warm by 5 K, cp = 4.18 kJ/(kg·K) gives a calculated minimum flow of about 4.02 L/min, assuming water density of 1 kg/L. The transfer surface must additionally transmit the 1.4 kW under actual temperature conditions. Acceptance therefore defines both utility supply at the connection and achieved function in the filled system.
Other utilities follow the same reasoning: electrical connection rating and typical energy consumption, required water quality and quantity, steam conditions where relevant, actuator air, and heat and liquids the room must accommodate. Details depend on the chosen configuration. The decisive question is whether these requirements can be met simultaneously under actual site conditions. Capable equipment connected to unsuitable building services cannot operate the process as expected.
Hypothetical acceptance case for minimum feed: A selected dosing assembly is tested at 2.0 mL/h over twelve hours. The substitute liquid has a measured density of 1.05 g/mL under the test conditions. Expected delivery is therefore 24.0 mL, or 25.2 g. An acceptance criterion defined solely for this teaching example permits cumulative delivery to differ from target by no more than 5%. The acceptable mass interval would be 23.94–26.46 g. This tolerance is not a general recommendation; an actual process must derive it from its sensitivity.
Measured delivery of 24.6 g differs by (24.6 − 25.2)/25.2 · 100 = −2.38%. If a justified measurement uncertainty assumed here as ±0.2 g is also considered, the resulting interval of 24.4–24.8 g lies entirely within the example acceptance interval. The cumulative test would pass under this decision rule. Agree how uncertainty will enter actual acceptance before testing; do not select its treatment after seeing the result.
This establishes only total delivery over twelve hours. Long interruptions followed by larger pulses could produce the same total. If the culture is sensitive to those fluctuations, the requirement also needs a temporal description of delivery and suitable evidence. Nor does the test yet establish adequate distribution after entry into the vessel. The example separates cumulative quantity accuracy, delivery over time and subsequent mixing. For repeatability, the dataset includes tubing dimensions, backpressure, delivery height, liquid condition, pump programme and the start and end of gravimetric measurement.
Include operation, data and lifetime costs
Access, assembly errors, sampling and maintenance time all affect daily use. Walk through the planned experiment from medium preparation to cleaning or disposal. Who lifts the vessel? Where do feed containers stand? Which connection must be changed during cultivation? What happens after a power or cooling failure?
Useful recording, time synchronisation, export and retrieval belong within the data requirements. User roles, change records and qualification documents depend on the actual application. A research experiment and regulated manufacturing do not automatically need the same documentation package.
A cost comparison includes purchase, consumables, energy, water, staff time, maintenance, downtime and disposal. Environmental assessment likewise needs a life-cycle boundary and local conditions. Neither visible polymer waste nor reusable steel alone determines the overall impact. Frequency of use and the resources consumed between cultures can alter the result substantially.
Describe data needs through an analysable test run. Include raw or sufficiently unaltered measured values, units, setpoints, actuator values, control states and events. Timestamps need a common reference. An export file becomes useful when it can be reopened with understandable channel identification and linked to samples. For long cultures, consider data gaps, restarts and resumption. Derive necessary roles and change records from intended use without assigning the same regulated scope to every research instrument.
Hypothetical TCO teaching calculation without market prices: Two fictional, technically suitable concepts, A and B, are compared over five years. All amounts are invented currency units, abbreviated CU. A requires an initial outlay of 120000 CU and fixed annual operating costs of 6000 CU; B requires 50000 and 3000 CU. Each fully budgeted run incurs an assumed 400 CU for A and 1000 CU for B. Residual value, financing, inflation and taxes are initially omitted to make the relationship clear.
| Cost component | A | B |
|---|---|---|
| Contact parts and consumables | 120 | 700 |
| Utilities | 80 | 40 |
| Labour, at fictional assigned cost | 160 | 200 |
| Disposal | 40 | 60 |
| Total | 400 | 1000 |
For n runs per year, the simplified model is TCO = I + 5 · Kfixed + 5 · n · Kvariable. At 40 annual runs, A totals 120000 + 30000 + 80000 = 230000 CU and B totals 50000 + 15000 + 200000 = 265000 CU. Across 200 runs, this is 1150 and 1325 CU per run, respectively. These are not prices or a general comparison between stainless steel and single-use. They show only the consequences of the disclosed assumptions.
At 20 annual runs, A totals 190000 CU and B totals 165000 CU, reversing the ranking. Calculated equality occurs at n = (120000 − 50000 + 5 · (6000 − 3000))/(5 · (1000 − 400)) = 28.33 runs per year. Whole runs and actual occupancy times then need consideration. A mathematically cheaper concept is not automatically available at the required capacity if its complete work cycle cannot accommodate the planned annual run count.
For an actual decision, replace every input with site-specific information and define the accounting boundary. When labour hours are monetised, explain whether they represent additional expenditure or committed existing capacity. Add failure and rejection assumptions only on a defensible basis. Without credible success probabilities, cost per successful batch is speculative. Environmental assessment needs its own resource and impact data; no environmental ranking follows from this financial teaching calculation.
Define supply boundaries and test responsibilities
A supply boundary describes which party provides which function up to which connection. It includes mechanical and electrical interfaces, gas, cooling, exhaust, drains and data. For each important connection, record interface type, operating conditions, capacity and responsibility. This prevents the same function from being included twice or supplied by neither party. External temperature-control units, gas supply, additional pumps and long-term retention of exported data are common sources of ambiguity.
The scope also includes prerequisites for performance testing. Who provides suitable substitute liquid, measuring instruments, simulated load and required site data? What cost and time windows are allocated to a trial with biological material? If a transfer result is promised only in water, record that explicitly. Later evaluation in culture medium requires its own scope and suitable criterion; it must not silently count as evidence already delivered.
A useful interface review follows a complete run. Identify every external service needed for preparation, inoculation, sampling, feeding, harvest and readiness for reuse. Include mobile containers, available floor space and operator access. For parallel operation, examine shared facilities. A vessel can perform well during cultivation yet miss the intended experimental throughput if preparation must wait for a shared resource already in use.
This makes the decision concrete: it concerns a defined combination of equipment, contact path, utilities, data functions and work sequence. Review changes to any component for their effects on the others. A larger pump may require another tubing route, which can change the smallest useful feed delivery or preparation. A sound choice therefore follows from a coherent operating and evidence concept, rather than a collection of independent maximum specifications.
Make the final decision verifiable
The final requirement list should distinguish mandatory criteria, desired reserve and unresolved development questions. Each major requirement needs a form of evidence: documentation, a transparent calculation, reference measurement or a trial using an appropriate liquid. The conditions under which that evidence was obtained are part of the claim.
For an uncertain process, flexible configuration and targeted experiments may be more useful than an apparently final commitment. For a well-understood process, repeatability, utilities and spare-parts planning often carry particular weight. Bioreactor types explains process geometries, aseptic processing the protection strategies, and scale-up the implications of later transfer.
The decision workflow starts with the process profile and mandatory criteria. Compare alternative configurations functionally, prioritise missing evidence and add site information. Give remaining candidates the same test cases and the same cost boundary. Once that information exists, assess weights, total cost and residual uncertainty together. The reasoned decision also states which changed assumptions would make another option sensible.
Build an acceptance evidence matrix: requirement identifier, agreed function, test, conditions, acceptance criterion, result and deviation. Factory Acceptance Testing, FAT, can examine appropriate functions at the supplier before shipment. Site Acceptance Testing, SAT, supplements checks after delivery under actual installed conditions. The terms do not imply automatically identical standard packages. Define the scope from the requirements and from what transport and installation could change.
A dosing test might include the intended tubing path, substitute liquid, duration and a gravimetric result. A cooling test must represent agreed fill, load and utility conditions. A data test creates a traceable sequence with a documented event and then checks export and retrieval. Define success before each test. Criteria should not subsequently shift merely to conceal an undesirable result; changes need a technical rationale.
At the site, pay particular attention to utilities, connection identification, actual assembly, room conditions and relevant data interfaces. A factory result can remain sufficient evidence only when the relevant function is unaffected by transport and installation and its reuse is justified. In regulated projects, acceptance, qualification and process validation belong within the applicable quality system. A successful FAT alone establishes neither the biological process nor its product quality.
Final handover includes the configuration actually delivered, agreed evidence, open points with assigned responsibility, operation and maintenance documentation, and data needed to repeat relevant tests. The objective is an operable, traceable system. If a central development question remains unresolved, subsequent use is bounded by an appropriate investigation plan. A signature on a component list does not replace establishing whether the intended process can demonstrably run with that configuration.
Frequently asked questions
Which material is the most sterile?
No material is inherently the most sterile. A suitably established sterile flow path, its integrity and aseptic operation of the whole process determine the outcome.
Can every bioreactor run both microbial and cell cultures?
Only if its mixing, gas, temperature and measurement ranges, together with the internals, suit both processes. Sharing a controller is not sufficient.
Is single-use always cheaper?
The answer depends on utilisation, run frequency, consumables, cleaning, utilities and staff time. A fair comparison considers the whole planned operation.
Terms in this chapter
49 terms explained in the glossary
- Acceptance criterion
- Antifoam
- Asepsis
- Autoclave
- Bioreactor
- Cell counting
- Chemically defined medium
- Clarification
- Cleaning
- Cleaning in place (CIP)
- Condensate
- Conformity decision and decision rule
- Culture medium
- Density
- Equipment maintenance
- Equipment occupancy time
- Extractables
- Factory acceptance test (FAT)
- Fed-batch culture
- Fermentation and fermenter
- Foam
- Gas holdup
- Harvest
- Headspace
- Heat transfer
- Leachables
- Measurement uncertainty
- Mixing time
- Nominal volume
- Operating envelope
- Operating expenditure (OPEX)
- Perfusion
- Product changeover
- Purity
- Qualification
- Scale-up
- Seal
- Site acceptance test (SAT)
- Sterilisation
- Sterilisation in place (SIP)
- Sterility
- Throughput
- Total Cost of Ownership (TCO)
- User requirements specification (URS)
- Validation
- Variable cost
- Viscosity
- Volumetric mass-transfer coefficient kLa
- Working volume