Glass laboratory bioreactors with stirrer drives, a supply unit and a separate touchscreen.
Real glass bioreactors in a laboratory setup: vessels, stirrer drives, pumps and the control unit work together as a system. Equipment example from bbi-biotech.Photo: bbi-biotech

Plan learning outcomes and evidence before equipment activities

Begin preparation with an observable achievement. After a module, a student might distinguish concentration from amount or narrow down a fault using two independent signals. Becoming familiar with a bioreactor is too vague to assess. A learning outcome specifies the task, necessary information and expected result. Only then choose whether a calculation, existing dataset, simulation or approved water run provides the appropriate learning environment.

Eight modules with assessable outcomes
ModuleLearning outcomeSubmission
1 DataDistinguish quantity, unit and statusCorrected data dictionary
2 Fed-batchBalance changing volumesAmount history and bounded interpretation
3 Water runDocument the measurement chain and test conditionsExperimental plan and analysed time series
4 ControlSeparate request, actual value and dynamicsSignal comparison with a testable hypothesis
5 BalanceDistinguish boundary and residualComponent balance with unresolved contributions
6 DesignSeparate effects from nuisance influencesJustified comparison plan
7 UncertaintyConnect a result with its limitationsUncertainty calculation and decision
8 ReportingSupport a traceable conclusionReport with reproducible calculations

The MIT course Chemical and Biological Reaction Engineering illustrates the connection between reaction understanding, transport and balances. The exercises here are original teaching proposals, rather than a reproduction of an MIT laboratory programme. Adapt their scope to prior knowledge and teaching time. An introductory course might complete the first two modules; an advanced course can connect them with model assessment and experimental design. Modules 1–2 require unit conversion and basic algebra; the model calculations in 3–4 use exponential functions, while Module 7 also requires basic measurement-uncertainty concepts.

Before group work, each student answers a short initial question independently. After discussion, assess the same concept again with different numbers. This reveals who understands the relationship and who has merely copied a spreadsheet formula. Explaining a mistake is itself a worthwhile technical achievement when the explanation restores the correct reasoning. Assessment should therefore reward evidence of understanding alongside a correct final number.

Module 1: Check units and data status

Exercise: A fictional export contains “feed 6”, “volume 1.5”, “DO 40” and “sample 14:20”. Construct a data dictionary before calculating anything. Each column needs a quantity, unit, reference condition, origin, meaning of time and data status. The correct initial response is that these four entries are insufficient. Explicitly identifying missing information is more useful than introducing an unnoticed assumption.

The lecturer supplies further information: feed is measured liquid flow of 6 mL min⁻¹, volume is 1.5 L of liquid, DO is a sensor reading expressed as a percentage of its documented saturation reference, and 14:20 is the availability time for a sample taken at 14:12. Initial conversions are now possible. Feed is 0.36 L h⁻¹, and dividing by current volume gives F/V = 0.24 h⁻¹. If volume grows, this ratio changes even when feed remains constant.

Interpretation: Do not convert the DO reading of 40 into mg L⁻¹ without the appropriate solubility reference and process conditions. Likewise, do not position the sample result at 14:20 on a process trend without showing its collection time. The eight-minute delay may matter substantially for later causal reasoning. Keep both timestamps as separate fields in the data dictionary.

A second entry gives 0.8 kW over 2.5 h. The result is 2.0 kWh of energy, rather than 2.0 kW of power. Ask students to carry units through every step to the final quantity. A spreadsheet can multiply numbers correctly while producing the wrong physical interpretation. The worked calculations provide further opportunities to practise this distinction. Dimensional consistency is necessary, although it cannot by itself establish that the chosen model represents the experiment.

A suitable submission contains the completed dictionary, two traceable conversions and three explicitly unresolved questions. Assess correct dimensions, separation of measurement from assumption, and treatment of time. For an extension, give two groups the same export with different decimal separators or missing status fields. Ask how an automated import should detect these differences before producing an apparently smooth trend. Missing observations initially remain missing; they should not silently become zero. Any subsequent interpolation needs a stated purpose and must retain access to the original observations.

Module 2: Balance a virtual fed-batch dataset

This module uses synthetic data only. It specifies neither an organism nor a procedure for running a culture. The model starts with 1.00 L in a vessel. For three hours, a virtual solution containing 20 g L⁻¹ of component S enters at a constant 0.10 L h⁻¹. The feed contains no X. The model has no withdrawal, evaporation or other volume change. X denotes an assumed biomass concentration. These figures do not represent the performance of a biological system.

Synthetic dataset for amount calculations
Time hVolume LX g L⁻¹S g L⁻¹
01.002.005.00
11.102.403.50
21.202.802.00
31.303.001.00

Exercise: Check the volume history, calculate vessel inventories X·V and S·V, and determine the amount of S that disappears over three hours. Plot concentration and amount separately. Axis labels must distinguish a quantity per litre from an inventory in the complete vessel. State explicitly which supplied values are assumed exact for this first exercise.

Solution: V = 1.00 + 0.10·t gives the tabulated volumes. X inventories are 2.00, 2.64, 3.36 and 3.90 g. S inventories are 5.00, 3.85, 2.40 and 1.30 g. Feed supplies 0.10·3·20 = 6.00 g of S. Initial inventory plus input minus final inventory gives net consumption of 5.00 + 6.00 − 1.30 = 9.70 g. Dividing the 1.90 g net increase in X by 9.70 g of disappearing S gives approximately 0.196 g g⁻¹.

This ratio is an observed net yield for the virtual interval. It establishes neither the fate of every element nor a metabolic mechanism. Product, gas and other component balances are among the missing information. The MIT lecture notes on biological reactors address the underlying relationship between feed, volume and biomass inventory. Their simplified examples also depend on stated modelling assumptions.

A follow-up question asks whether constant X = 2 g L⁻¹ would mean no biomass formation when volume increases. It would not: increasing volume from 1.00 to 1.30 L raises inventory from 2.00 to 2.60 g. Concentration can stay constant while amount grows. Explaining this is the module's most important evidence of learning. Introduce sampling only in a new model version, including both withdrawn volume and the associated component amount. This prevents an added complication from silently changing the original calculation boundary.

Module 3: Obtain useful evidence from a technical water run

A water run without a culture can teach measurement, timing and equipment understanding. It is appropriate practical work only when the specific equipment, activity and supervision have been approved. The vessel, drive, electrical components and any temperature-control equipment retain engineering hazards without a biological culture. Local instruction establishes operating authority, permitted conditions and stop criteria. The page on engineering safety supports preparation of the relevant questions.

Planning exercise: Before practical work, define the property to be observed, such as the time response of an existing measurement channel. Record sensor location, acquisition interval, event marker, initial condition and the intended conclusion. Do not derive operating changes or addition of a substance from a general teaching webpage; these must belong to the locally approved exercise. A recorded water-run time series provides an alternative when practical work is unavailable.

A useful analysis compares the raw signal with the documented event. Students first identify the onset of change, response speed and whether a stable final region is visible. Distinguish transport delay from a slow sensor response where the available evidence permits. One channel usually cannot separate both contributions uniquely. State this limitation explicitly rather than labelling every delay as poor mixing. The answer can be technically strong even when it identifies that further information is required.

For a purely virtual ideal well-mixed vessel at constant volume 2 L with equal inflow and outflow of 0.5 L min⁻¹, V/F is 4 min. For a component that neither reacts nor undergoes other losses in the model, the normalised outlet following an ideal inlet concentration step would reach 1 − e⁻¹ = 63.2% of the complete change after that time. This is a model prediction, not a specified mixing time for real equipment. Sensor lag, pipe volume and incomplete mixing can modify the measured response.

The MIT notes on non-ideal reactor mixing distinguish ideal residence-time models from real departures. Submissions should therefore separate model assumptions, observed behaviour and proposed causes. A water run establishes neither biological performance nor sterility or suitability for every other liquid. Strong work proposes an appropriate additional observation while respecting the requirement for approval of any further practical activity.

Module 4: Explain signals and controller responses

Control becomes clearer when the requested action and actual effect are plotted separately. Simulated or recorded data are sufficient for this module. Required channels are setpoint, measured controlled variable, controller output, actual manipulated variable, operating mode and events on a common time base. A colourful process trend without these supporting channels is often insufficient for interpretation.

Hypothetical exercise: Two virtual systems start at the same controlled-variable value. After a disturbance, System A requests increasing actuator effort. Actual flow follows until a documented limit and then remains constant; the controlled variable does not regain setpoint. System B shows a similar output request, but actual flow remains unchanged well below the documented limit. What explanation distinguishes the two cases?

Interpretation: A is consistent with a genuinely reached actuator limit. For B, first establish whether the requested change occurred at all or whether the actual-flow measurement is faulty. Controller output alone establishes neither delivered flow nor remaining process capacity. A useful additional observation would independently confirm actual flow. For each hypothesis, students specify an observation that would support it and one that would weaken it. This turns a plausible story into a testable explanation.

The calculation uses an ideal first-order measurement channel. Following a true step from 20 to 30 units, with an assumed time constant of 30 s, its signal is y(t) = 30 − 10·e^(−t/(30 s)). After 30 s the reading is 26.32 units; after 90 s it is 29.50. A slow measurement therefore does not prove an equally slow process. The equation contains neither additional dead time nor noise. Add these separately in a later simulation so that their effects remain distinguishable.

Another discussion concerns staged use of actuators versus true cascade control. An outer controller successively using several actuators does not automatically create an inner closed feedback loop. Ask students to mark which measured variable is returned to each controller. The page on measurement and control develops this architectural distinction and explains why names in a software interface are insufficient to establish the actual control structure.

Assess correct signal assignment and a justified next diagnostic step, rather than the most aggressive tuning proposal. Changes to controller parameters, limits or protective functions are not unsupervised exercises on laboratory equipment. Simulation can reveal saturation and delayed recovery without altering actual protective boundaries. Captions and conclusions must retain the distinction between a simulated response and an observed equipment result.

Module 5: Investigate an incomplete balance

A balance is both a calculation and an agreement about scope. Before inserting numbers, define the substance or element, interval and physical boundary. The general relationship is: initial inventory plus input minus output plus net formation equals final inventory. For an elemental balance without nuclear transformation, chemical formation of the element is absent. A component balance can nevertheless depend strongly on chemical reaction.

Hypothetical exercise: A completely virtual process receives 10.0 C-mol of carbon during a defined interval. Initial and final inventories have already been assessed separately, and this teaching case has no additional net accumulation. Recorded product streams contain 3.0 C-mol, a biomass fraction contains 2.0, gas contains 4.0, and other dissolved substances contain 0.5. Here C-mol means one mole of carbon atoms, not one mole of arbitrary molecules. Calculate recovery and propose two different explanations for the residual.

Solution: The sum is 9.5 C-mol, giving 95% recovery. The positive residual is 0.5 C-mol, or 5% of the input. It does not establish a particular unknown compound. Possibilities include an unrecorded stream, incorrect carbon assignment, measurement error or an inventory change overlooked despite the assumption. A defensible conclusion identifies the discrepancy without automatically assigning its mechanism.

An extension examines a unit error. One mole of a compound containing six carbon atoms contains six C-mol. Adding molecular amounts of different compounds without this conversion can create apparent failure of conservation. Similarly, grams of biomass cannot be converted into C-mol without a justified composition. The module therefore requires a small conversion table rather than one opaque spreadsheet sum. Every conversion must identify its input basis and the assumption that makes it possible.

The group then chooses which additional information would be most valuable. Another measurement of a small, well-characterised contribution will not automatically resolve a large uncertain gas balance. Prioritise a contribution capable of materially influencing the residual and whose measurement can actually be improved. Distinguish numerical closure from independent confirmation: adjusting a free correction factor until recovery becomes exactly 100% does not establish the underlying explanation.

The submission contains a flow diagram with its boundary, a fully labelled table, the balance residual and a justified follow-up question. The same reasoning is required in process development when an improved product result accompanies other losses or a changed harvest volume. A class can explore this relationship without carrying out a biological reaction. A useful final sentence states which conclusion remains possible even while the discrepancy is unresolved.

Module 6: Design a comparison before collecting results

Experimental design starts with the statement a comparison should support. Demonstrating a difference, screening important factors and locating a favourable region need not require the same design. The NIST guidance on selecting an experimental design distinguishes these objectives explicitly. In teaching, a justified choice of response variable is already a worthwhile task in its own right.

Virtual exercise: Two abstract factors, A and B, each have a low and a high level. The response is a dimensionless simulation metric. At low A/low B it is 10; at high A/low B it is 14; at low A/high B it is 12; and at high A/high B it is 20. Calculate simple differences before opening regression software. These levels are not actual operating settings or instructions for changing equipment.

Complete combinations for the virtual two-factor case
ABResponse
lowlow10
highlow14
lowhigh12
highhigh20

Solution: The A difference is four units at low B and eight at high B: its effect depends on B. The difference between these differences is four. Averaging over both B levels gives an A effect of six; the corresponding average B effect is four. Studying A only at low B would miss its larger effect at high B. These findings describe the synthetic dataset. Without independent replication, they do not establish a statistically supported effect in an actual process.

A planned practical comparison also needs to address time. If all low settings are observed in the morning and high settings in the afternoon, a change through the day can be confused with the factor effect. Students propose appropriate replication, meaningful randomisation and treatment of known nuisance influences. With two experimental days, each day can contain the relevant comparisons instead of assigning an entire treatment to its own day.

The NIST explanation of randomised block designs describes handling of known nuisance factors. Blocking does not eliminate every possible interaction. Randomisation also cannot guarantee perfect balance in a small sample. Students should therefore justify the design and identify the limitations that remain. A carefully bounded comparison is more useful than a broad claim unsupported by the chosen experimental units.

The submission specifies response, factor levels, experimental unit, repetitions, order and analysis. Repeated analyses of one sample are not independent repetitions of the entire process. Recognising this distinction is a major learning outcome. An extension asks why four corner points cannot establish the response between levels or identify an optimum outside the investigated region. These questions connect a simple classroom table with the real limits of extrapolation.

Module 7: Translate uncertainty into a defensible statement

Uncertainty is easiest to understand when connected to a decision. A value with many decimal places does not establish how well the underlying quantity is known. Students first distinguish resolution, repeatability, calibration contributions and possible systematic influences. Repeatedly reading one stable display does not automatically capture uncertainty in the complete measurement procedure.

Hypothetical exercise: Determine a component amount from M = C·V. Concentration C is 2.00 g L⁻¹ with standard uncertainty u(C) = 0.06 g L⁻¹, and volume V is 1.50 L with standard uncertainty u(V) = 0.015 L. Assume independent contributions and suitable linear uncertainty propagation for this teaching model. Calculate M and its combined standard uncertainty. The supplied uncertainties are already standard uncertainties; do not arbitrarily divide them by another factor.

Solution: M is 3.00 g. Relative contributions are 0.03 and 0.01, giving u(M)/M = √(0.03² + 0.01²) = 0.0316 and u(M) = 0.0949 g. With an assumed coverage factor k = 2, expanded uncertainty U is 0.190 g, appropriately rounded to approximately 3.00 g ± 0.19 g. A particular coverage probability requires further assumptions; the number k = 2 alone does not create an exact demonstration of 95% coverage.

The NIST treatment of uncertainty propagation addresses the influence of the model and covariance. The exercise therefore asks what changes if C and V come from measurements with shared influences. Independence then needs assessment, and a simple sum of squares may be insufficient. An additional common calibration contribution also does not disappear through unlimited repetition of the same measurement arrangement.

For the decision exercise, suppose a hypothetical lower limit is 2.90 g. The numerical estimate 3.00 g is above it, but the displayed uncertainty interval extends below it. Without a specified decision rule, this does not support an unambiguous conformity statement. Students identify the rule and false-decision risk that the responsible authority would need to establish. The task is not permission to select a limit retrospectively to suit a preferred outcome.

Assess arithmetic, assumptions and the wording of the conclusion together. A strong answer identifies information that would improve the decision, such as better concentration measurement because its relative contribution dominates here. It does not suggest that every extra decimal place increases confidence. This exercise can be completed entirely on a computer and also works as a structured review of another group's laboratory report.

Complete uncertainty budget and paired experimental comparison.

Module 8: Write a laboratory report that can be checked

A laboratory report should let another person assess a technical statement. It needs more than a chronological list of activities. Start with the question, defined boundary and principal result, then show the data and calculations supporting that result, deviations encountered and limitations affecting interpretation. An unresolved outcome is acceptable when the reasoning behind it is clear.

Writing exercise: Select one of the preceding cases and prepare a short result section with a table or graph. Another group should be able to reproduce the central number from the supplied information. They may ask questions, but should not need to guess missing units or modelling assumptions. Preserve a traceable connection between raw data, processing steps and the final figure. A reader should be able to identify which version of the data produced the reported result.

Example assessment rubric totalling 100 points
AreaPointsObservable evidence
Question and boundary15Response, interval and system clearly defined
Data and method20Origin, units, timing and processing visible
Calculation and presentation25Reproducible result, appropriate axes and rounding
Interpretation25Observation, hypothesis and limitation distinguished
Reflection and next step15Useful follow-up question and justified improvement

This rubric is a teaching proposal, not an examination regulation. Institutional requirements for participation, safety, independent work and permitted aids also apply. Safe conduct is not a minor item that can simply be offset by strong calculation marks. Stopping an activity for a justified safety reason can demonstrate sound technical judgement. A replacement recorded dataset can then support the analytical part of the exercise.

Peer review asks specific questions: which quantity was measured, calculated or assumed? Does a result refer to sample collection time? Are excluded observations accompanied by a visible reason? Is a smoothed curve labelled as processed data? Could another explanation produce the same observations? These questions challenge the statement without attacking its author. Feedback should identify the relevant passage and propose an actionable improvement.

A weak sentence might claim that the experiment proves optimal control. A stronger version states that, in the simulated disturbance analysed, actual flow followed the request up to the specified limit and a control error then remained. This identifies the subject and observation without an unsupported general claim. Negative findings also need precision: failing to detect a difference does not prove equality when the measurement may be insufficiently precise for the difference of interest.

When using spreadsheets, scripts or digital assistance, responsibility for technical checking remains visible. Automatically generated prose does not replace verification of numbers and references. Declare aids as required by institutional rules and retain the underlying calculation. A brief oral transfer exercise using changed inputs can reveal whether the student understands the statement rather than merely reproducing its wording.

Combine the modules into a coherent laboratory course

The modules can form one evolving case record. Students first create a data dictionary, then an amount calculation, a diagnosis and a justified proposal for the next experiment. Each new file has an identifiable version and origin. Documentation becomes a working input to the next decision rather than an administrative task added after the analysis.

Possible group roles include data checking, modelling, independent calculation and presentation. Rotate these between modules so that one person does not always calculate while another only prepares slides. Before submission, each member explains a selected relationship. Where prior knowledge differs, retain a common core question and offer additional model variants for deeper work without making the main task unnecessarily difficult to navigate.

The lecturer should prepare more than an ideal dataset. A second version can contain a missing unit, delayed timestamp or inconsistent balance. These are deliberately created teaching errors and should be disclosed afterwards. They are not permission to alter actual laboratory records. A useful closing discussion explains how the error was detected, which results it affected and which conclusions remained valid. This makes correction part of competent work rather than something to conceal.

Practical participation and analytical achievement should be connected without making one unnecessarily dependent on the other. If equipment is unavailable for approved use or a measurement must be stopped, the same analytical question can be addressed with existing data. Clearly identify the difference between personally acquired and supplied observations. This preserves both the learning outcome and honest reporting. Students can still explain what an appropriate acquisition plan would have required.

The page on reading and further learning provides additional books, courses and reference material. This chapter complements it with concrete exercises and assessment criteria. Successful learning is visible when students encounter a new number or curve and ask what was measured, which boundary applies and what decision the information can support, before accepting it as a finished conclusion.

Frequently asked questions

Does a bioreactor class always need a living culture?

No. Units, balances, signal interpretation, experimental design and uncertainty can be studied with synthetic or existing data. A technical water run can add practical competence, but requires approval and supervision for the specific laboratory activity.

How can group work be assessed fairly?

A transparent rubric assesses the shared result, while short individual explanations or transfer questions reveal personal understanding. Rotating roles and recording contributions prevent the same people from always doing only calculations or presentations.

Terms in this chapter

36 terms explained in the glossary